Proline Promag H 300
Endress+Hauser 17
Plain text display With information on cause and remedial measures
Additional information on remote operation → 82
Web server
Plain text display With information on cause and remedial measures
Light emitting diodes (LED)
Status information Status indicated by various light emitting diodes
The following information is displayed depending on the device version:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
Ex connection data Safety-related values
Order code for
"Output; input 1"
Output type Safety-related values
"Output; input 1"
26 (+) 27 (–)
Option BA Current output 4 to 20
mA HART
U
nom
= 30 V
U
max
= 250 V
Option GA PROFIBUS PA U
nom
= 32 V
U
max
= 250 V
Option MA Modbus RS485 U
nom
= 30 V
U
max
= 250 V
Option SA FOUNDATION Fieldbus U
nom
= 32 V
U
max
= 250 V
Order code for
"Output; input 2";
"Output; input 3"
Output type Safety-related values
Output; input 2 Output; input 3
24 (+) 25 (–) 22 (+) 23 (–)
Option B Current output 4 to 20
mA
U
nom
= 30 V
U
max
= 250 V
Option D User configurable input/
output
U
nom
= 30 V
U
max
= 250 V
Option E Pulse/frequency/switch
output
U
nom
= 30 V
U
max
= 250 V
Option F Double pulse output U
nom
= 30 V
U
max
= 250 V
Option H Relay output U
nom
= 30 V
I
nom
= 100 mA DC/500 mA AC
U
max
= 250 V
Option I Current input 4 to 20 mA U
nom
= 30 V
U
max
= 250 V
Option J Status input U
nom
= 30 V
U
max
= 250 V