Proline Promag H 300
Endress+Hauser 21
FOUNDATION Fieldbus
Manufacturer ID 0x452B48
Ident number 0x103C
Device revision 1
DD revision Information and files under:
• www.endress.com
• www.fieldbus.org
CFF revision
Interoperability Test Kit (ITK) Version 6.1.2
ITK Test Campaign Number Information:
• www.endress.com
• www.fieldbus.org
Link Master capability (LAS) Yes
Choice of "Link Master" and
"Basic Device"
Yes
Factory setting: Basic Device
Node address Factory setting: 247 (0xF7)
Supported functions The following methods are supported:
• Restart
• ENP Restart
• Diagnostic
Virtual Communication Relationships (VCRs)
Number of VCRs 44
Number of link objects in VFD 50
Permanent entries 1
Client VCRs 0
Server VCRs 10
Source VCRs 43
Sink VCRs 0
Subscriber VCRs 43
Publisher VCRs 43
Device Link Capabilities
Slot time 4
Min. delay between PDU 8
Max. response delay 20
Transducer Blocks
Block Contents Output values
Setup Transducer Block
(TRDSUP)
All parameters for standard commissioning. No output values
Advanced Setup
Transducer Block
(TRDASUP)
All parameters for more accurate measurement
configuration.
No output values
Display Transducer
Block
(TRDDISP)
Parameters for configuring the local display. No output values
HistoROM Transducer
Block
(TRDHROM)
Parameters for using the HistoROM function. No output values