Proline Promag H 300
60 Endress+Hauser
Coupling SC DIN 11851, threaded adapter
1.4404 (316L)
Order code for "Process connection", option DCS
DN
[mm]
Suitable for pipe EN 10357 (DIN 11850)
[mm]
A
[mm/in]
B
[mm]
L
[mm]
2 to 8 12 × 1 (DN 10) Rd 28 × ¹⁄₈ 10 174
15 18 × 1.5 (ODT ¾") Rd 34 × ¹⁄₈ 16 174
25 28 × 1 or 28×1.5 Rd 52 × ¹⁄₆ 26 190
40 42 × 1.5 Rd 65 × 1/6 38 260
50 54 × 1.5 Rd 78 × 1/6 50 260
65 70 × 2 Rd 95 × 1/6 66 270
80 85 × 2 Rd 110 × 1/4 81 280
100 104 × 2 Rd 130 × 1/4 100 290
125 129 × 2 Rd 160 × 1/4 125 380
150 154 × 2 Rd 160 × 1/4 150 390
Surface roughness: R
a
≤ 0.8 µm, optional ≤ 0.38 µm
Please note the internal diameters of the measuring tube and process connection (B) when cleaning with pigs.
Coupling DIN 11864-1, aseptic threaded hygienic connection, Form A
1.4404 (316L)
Order code for "Process connection", option DDS
DN
[mm]
Suitable for pipe EN 10357 (DIN 11850)
[mm]
A
[mm/in]
B
[mm]
L
[mm]
2 to 8 Pipe 13 × 1.5 (DN 10) Rd 28 × ¹⁄₈ 10 170
15 Pipe 19 × 1.5 Rd 34 × ¹⁄₈ 16 170
25 Pipe 29 × 1.5 Rd 52 × ¹⁄₆ 26 184
40 42 × 1.5 Rd 65 × 1/6 38 256
50 54 × 1.5 Rd 78 × 1/6 50 256
65 70 × 2 Rd 95 × 1/6 66 266
80 85 × 2 Rd 110 × 1/4 81 276
100 104 × 2 Rd 130 × 1/4 100 286
Surface roughness: R
a
≤ 0.8 µm, optional ≤ 0.38 µm
Please note the internal diameters of the measuring tube and process connection (B) when cleaning with pigs.
Coupling ISO 2853, threaded adapter
1.4404 (316L)
Order code for "Process connection", option ICS
DN
[mm]
Suitable for pipe ISO 2037
[mm]
DN
Clamp ISO 2853
[mm]
A
[mm/in]
B
[mm]
L
[mm]
40 38 × 1.6 38 Tr 50.5 × 3.175 35.6 256
50 51 × 1.6 51 Tr 64 × 3.175 48.6 256
65 63.5 × 1.6 63.5 Tr 77.5 × 3.175 60.3 266
80 76.1 × 1.6 76.1 Tr 91 × 3.175 72.9 276
100 101.6 × 2 101.6 Tr 118 × 3.175 97.6 286
Surface roughness: R
a
≤ 0.8 µm, optional ≤ 0.38 µm
Please note the internal diameters of the measuring tube and process connection (B) when cleaning with pigs.