Technical data Proline Promass F 100 Modbus RS485
104
DN [% o.r./bar] [% o.r./psi]
[mm] [in]
50 2 –0.008 –0.0006
80 3 –0.009 –0.0006
100 4 –0.007 –0.0005
150 6 –0.009 –0.0006
250 10 –0.009 –0.0006
Design fundamentals o.r. = of reading, o.f.s. = of full scale value
BaseAccu = base accuracy in % o.r., BaseRepeat = base repeatability in % o.r.
MeasValue = measured value; ZeroPoint = zero point stability
Calculation of the maximum measured error as a function of the flow rate
Flow rate Maximum measured error in % o.r.
ZeroPoint
BaseAccu
100
³
A0021332
± BaseAccu
A0021339
ZeroPoint
BaseAccu
100
<
A0021333
ZeroPoint
MeasValue
100
±
A0021334
Calculation of the maximum repeatability as a function of the flow rate
Flow rate Maximum repeatability in % o.r.
½ ZeroPoint
BaseRepeat
100
³
A0021335
A0021340
½ ZeroPoint
BaseRepeat
100
<
A0021336
ZeroPoint
MeasValue
100
½ ±
A0021337
Example for maximum measured error
E [%]
0 10 20 30 40 50 60
0
0.5
1.0
1.5
2.0
2.5
70 80 90 100
Q [%]
A0028808
E Maximum measured error in % o.r. (example with PremiumCal)
Q Flow rate in % of maximum full scale value