EasyManua.ls Logo

Endress+Hauser Proline Promass F 100 - Page 63

Endress+Hauser Proline Promass F 100
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promass F 100 Modbus RS485 Commissioning
63
Navigation
"Setup" menu → Advanced setup → Calculated values
Calculated values
Corrected volume flow calculation
Corrected volume flow calculation
→  63
External reference density
→  63
Fixed reference density
→  63
Reference temperature
→  63
Linear expansion coefficient
→  64
Square expansion coefficient
→  64
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Corrected volume flow calculation Select reference density for
calculating the corrected
volume flow.
Fixed reference
density
Calculated
reference density
Reference density
by API table 53
External reference
density
Calculated reference
density
External reference density In the Corrected volume flow
calculation parameter, the
External reference density
option is selected.
Shows external reference
density.
Floating point
number with sign
Fixed reference density The Fixed reference density
option is selected in the
Corrected volume flow
calculation parameter
parameter.
Enter fixed value for reference
density.
Positive floating-
point number
1 kg/Nl
Reference temperature The Calculated reference
density option is selected in
the Corrected volume flow
calculation parameter.
Enter reference temperature
for calculating the reference
density.
–273.15 to 99 999 °C Country-specific:
+20 °C
+68 °F