EasyManua.ls Logo

Endress+Hauser Proline Promass P 300 - Clamp Connections; Tri-Clamp

Endress+Hauser Proline Promass P 300
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promass P 300
64 Endress+Hauser
Clamp connections
Tri-Clamp
L
A
B
(1")DN 25DN 8, 15 (¾", ½")
A
B
A0023342
 37 Engineering unit mm (in)
Length tolerance for dimension L in mm:
+1.5 / –2.0
Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FHW
DN
[mm]
Clamp
[in]
A
[mm]
B
[mm]
L
[mm]
8 ½ 25.0 9.40 362
15 ¾ 25.0 15.75 466
25 1
1)
50.4 22.10 606
40
1)
50.4 34.80 731
50 2
1)
63.9 47.50 853
Ra
max
0.76 µm: order code for "Measuring tube material", option BB, TD
Ra
max
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
1) The connection complies with the hygienic clamp dimensions as per ASME BPE.
½" Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FCW
DN
[mm]
Clamp
[in]
A
[mm]
B
[mm]
L
[mm]
15 ½ 25.0 9.40 466
Ra
max
0.76 µm: order code for "Measuring tube material", option BB, TD
Ra
max
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
¾" Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FFW
DN
[mm]
Clamp
[in]
A
[mm]
B
[mm]
L
[mm]
8 ¾ 25.0 15.75 362
Ra
max
0.76 µm: order code for "Measuring tube material", option BB, TD
Ra
max
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG

Table of Contents