Installation Proline t-mass 65
10 Endress+Hauser
Aligning the calculated insertion depth
Aligning the insertion version to flow direction
Check and ensure that the sensor is aligned vertically at a 90° angle on the pipe/duct.
Turn the sensor so that the arrow marking matches the direction of flow.
A0011848
a. Tighten the upper nut of compression fitting such that
the sensor can still be adjusted.
b. Align the scale to the calculated insertion depth.
A0007536
A0007537
a
b
230
220
210
200
190
180
9
8
7
90° (±3°)
90°(±7°)