Appendix RIA45
72 Endress+Hauser
Description Delay for switching the relay.
User entry 0-9999
Factory setting 0
Operating mode
Navigation
Expert → Output → Relay 1/Relay 2 → Operating mode
Description Normally closed = breaker contact
Normally opened = maker contact
Options normally closed
normally opened
Factory setting normally closed
Failure mode
Navigation
Expert → Output → Relay 1/Relay 2 → Failure mode
Description Normally closed = breaker contact
Normally opened = maker contact
Options normally closed
normally opened
Factory setting normally closed
Untermenü "Application"
Navigation
Expert → Application
Submenu "Calc value 1"/"Calc value 2"
Navigation
Expert → Application → Calc value 1/Calc value 2
Description Settings for the mathematics channels.
Additional information The following parameters are available for math 1 and math 2.
Decimal places
Navigation
Expert → Application → Calc value 1/Calc value 2 → Decimal places
Description Configures the number of decimal places for the display
Options XXXXX
XXXX.X
XXX.XX
XX.XXX
X.XXXX
Factory setting XXX.XX
Failure mode
Navigation
Expert → Application → Calc value 1/Calc value 2 → Failure mode

Other manuals for Endress+Hauser ria45