SS2100 Analyzer
34 4900002253 rev. C 9-30-20
SpectraSensors recommends using 1/4” O.D x 0.035” wall thickness, seamless
stainless steel tubing. Refer to the system layout drawings for supply and
return port locations.
Toconnectthesamplesupplyline
1. First confirm that the sample probe is correctly installed at the
process supply tap and that the sample probe isolation valve is
closed.Also, confirm that the field pressure reducing station is
installed properly at the sample probe and that the pressure
regulator at the field pressure reducing station is closed (adjustment
knob turned fully counterclockwise).
2. Check that the relief valve vent line is properly installed from the
field pressure reducing station to the low pressure flare or
atmospheric vent connection.
3. Determine appropriate tubing route from the field pressure reducing
station to the SCS.
4. Run stainless steel tubing from the field pressure reducing station
set for the specified supply pressure to the sample supply port of the
SCS.
5. Bend tubing using industrial grade benders, check tubing fit to
ensure proper seating between the tubing and fittings.
Process samples may contain hazardous material in potentially
flammable and/or toxic concentrations. Personnel should have a
thorough knowledge and understanding of the physical properties
and safety precautions for the sample contents before installing
the SCS.
Consult sample probe manufacturer instructions for proper
installation procedures.
The process sample at the sample tap may be at a high pressure.
Use extreme caution when operating the sample probe isolation
valve and field sample reducing pressure regulator.
All valves, regulators, switches, etc. should be operated in
accordance with site lock-out/tag-out procedures.