SS2100 Analyzer
58 4900002253 rev. C 9-30-20
4. Tighten all new fittings 1-1/4 turns with a wrench from finger tight.
For connections with previously swaged ferrules, thread the nut to
the previously pulled up position, then tighten slightly with a wrench.
5. Reset the scrubber lifetime monitor with the New Scrub Installed
parameter and the General Fault Alarm with the Reset option for
the General Alarm
DO parameter (see “To change parameters in
Mode 2” in the Firmware Manual for your analyzer).
6. Restart the SCS.
7. Check all connections for gas leaks. SpectraSensors recommends
using a liquid leak detector.
8. Re-validate the system with an appropriate gas standard following
the instructions under “Validating the Analyzer” in the Firmware
Manual for your analyzer.
9. Purge the scrubber and scrubber efficiency indicator assembly with
nitrogen to remove all flammable gas and cap the inlet and outlet.
DisposalofUsedScrubbers
1. Discard used scrubber and scrubber indicator in an appropriate leak-
proof receptacle.
Service Contact
For Technical Service, refer to our website for the list of local sales channels in
your area (https://www.spectrasensors.com/contact).
H
2
S scrubbers and scrubber indicators contain Copper (II) Oxide
[CAS# 1317-38-0] and basic cupric carbonate [CAS# 12069-69-
1], which are harmful if swallowed and toxic to aquatic organisms.
Handle with care and avoid contact with the internal substances.
Depleted H
2
S scrubbers and scrubber indicators contain
predominantly Copper (II) Sulfide [CAS# 1317-40-4] with some
remaining Copper (II) Oxide [CAS# 1317-38-0] and basic cupric
carbonate [CAS# 12069-69-1], each of which are odorless dark
powders that require few special precautions other than avoiding
contact with the internal substances, keeping the scrubber tightly
sealed and protecting the contents against humidity.

Table of Contents

Other manuals for Endress+Hauser SpectraSensors SS2100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser SpectraSensors SS2100 and is the answer not in the manual?

Endress+Hauser SpectraSensors SS2100 Specifications

General IconGeneral
BrandEndress+Hauser
ModelSpectraSensors SS2100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals