SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 31
Fig.
SOV
Description
Terminal
Relay Rating Ith
Fig 17
S1
Scrubber Solenoid
1
6A
2
S2
Main/Val Solenoid
3
4
Fig 18
S1
Scrubber Solenoid
1
2
Fig 19
S1
Scrubber Solenoid
1
2
S2
Val 1 Solenoid
5
6
Fig 20
S1
Scrubber Solenoid
1
2
S2
Main/Val Solenoid
3
4
S3
Val 1 Solenoid
5
6
S4
Val 2 Solenoid
7
8
Terminal block (X2) solenoid valve connections
NOTICE
For valve configurations refer to Fig 17 through Fig 20.
5. Verify that each connection is secure.
6. Close the analyzer enclosure cover.
4.11 Application of gland lubricant
To ensure proper installation, Endress+Hauser recommends using STL8 screw thread lubricant or equivalent on all
conduit screw thread and its taped opening.
STL8 Screw Thread Lubricant is a lithium based, anti-galling substance with excellent adhesion that maintains rain-
tightness and grounding continuity between conduit fittings. This lubricant has proven very effective between parts
made of dissimilar metals, and is stable in temperatures from 20 F to 300 F.
NOTICE
Do not use this lubricant on exposed current-carrying parts.
1. Holding the fitting piece at one end, generously apply the lubricant on the male threaded surface (at least five
threads wide) as shown below.