SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 39
5 Specifications
See analyzer calibration report
See analyzer calibration report
Typically less than 20 seconds
Environmental temperature range
–20 C to 50 C (–4 F to 122 F) — Standard
–30 C to 60 C (–14 F to 140 F) — Optional
Heated enclosure temperature
Shipment and storage temperatures
≥ –20 °C (–4 °F) — Standard
> 0 °C (32 °F) — Analyzers with perm tube validation
Environmental relative humidity
5% to 95%, Non-condensing
Sample cell operating pressure range
800 to 1200 mbar — Standard; 950 to 1700 mbar — Optional
70 kPaG (10 PSIG) max to spectrometer cell
0.5 to 4 SLPM (0.02 to 0.1 SCFM)
1
None for gas phase glycol, methanol, amines, or mercaptans
Electrical & communications
120 or 240 VAC ± 10%, 50 to 60 Hz standard; ~300 W
1
Analog: Isolated Analog channels, 1200 ohms at 24 VDC max
Outputs: Qty (2) 4 to 20 mA (measurement value)
Input: Qty (1) 4 to 20 mA (pipeline pressure)
1
Serial: Ethernet and RS-485 half-duplex — Standard
Outputs: (5); Hi/Lo Alarm, General Fault, Validation Fail
1
, Validation 1 Active
1
,
Validation 2 Active
1
Inputs: (2); Flow Alarm
1
, Validation Request
1
Modbus Gould RTU, Daniel RTU or ASCII
Diagnostic value examples
Detector Power (Optics Health), Spectrum Reference Comparison and Peak Tracking
(Spectrum Quality), Cell Pressure and Temperature (Overall System Health)
Concentration, Cell Pressure, Cell Temperature, and Diagnostics
IP66 Copper-free Aluminum with Weather-resistant Polyurethane Powder Coating, 80 to
120 micron thickness
Sample system enclosure(s)
IP55 (min.) 304 or 316L Stainless Steel
670 mm H ´ 1270 mm W ´ 248 mm D (26.3 in H x 50 in W x 9.7 in D)
Approximately 145 kg (320 lbs) (not including Sample Conditioning System)
316L Series Polished Stainless Steel — Standard
Supply voltage not to exceed +10% of nominal. Transient over-voltages according to Over-voltage category II.
Dimensions are for analyzer only. Consult system drawings for analyzers with sample conditioning systems.