SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 35
Fig 23. Optically isolated RS-232-to-RS-485 converter DIP switches
A. Internal termination not used. 9600 Baud, RS-245 two-wire
B. Cable shield should be grounded.
RS-485 2-Wire
Half Duplex
120 Ω Built-in
Termination
External or no
Termination
Output signal connections (two-wire RS-485 configuration)
Hazardous voltage and risk of electric shock. Turn off and lock out system power before opening the electronics
enclosure and making any connections.
Certified compound barrier seal type glands and cables shall be used where appropriate in compliance with local
regulations.
See Fig 7. Components in SS2100i-2 electronics enclosure → for locating fuses. If you need to replace a fuse,
use only the same type and rating of fuse as the original as shown in the following fuse specifications tables. To
re-order parts, refer to Analyzer parts → .
Time-out selections are equal to one character time at the indicated baud rate.
To achieve this time-out, an appropriate through-hole resistor must be placed in the R11 location on the converter PCB.