10 Messages
54 / 70
80114-770_V23802
2021-01
APC-A-151 Argon gas (Ar) Argon gas is required for the selected program, but no argon gas
pressure is detected.
Connect the gas hose at the rear of the APC 3 argon plasma module.
S-A-152 APC instrument Automatic argon gas filling of the instrument has been started.
S-A-166 APC 3 argon plasma module The argon plasma module APC 3 is not ready for operation.
Open the valve of the argon gas bottle.
Ensure that the gas hose and the sensor cable of the pressure regu-
lator are connected at the rear of the APC 3 argon plasma module.