11 General Technical Data
56 / 70
80114-770_V23802
2021-01
Residual pressure display Manometer on the gas bottle
The APC 3 switches off at an input pressure of < 3 x 10
5
Pa < 3 bar < 43.5 psi
Dimensions and weight
Width x height x depth 415 x 100 x 375 mm / 16.3“ x 3.9“ x 14.8“
Weight 5.3 kg / 11.7 lbs
Ambient conditions for transport and storage of unit
Temp erature -29 °C to +60 °C / -20 °F to +140 °F
Relative humidity 15%–85%
Ambient conditions for operation of the unit
Temp erature +10 °C to +40 °C / +50 °F to +104 °F
Relative humidity 15% - 80%, non-condensing
Air pressure 54 kPa – 106 kPa / 7.8 psi - 15.4 psi
Maximum operating altitude 5000 m / 16404 ft above sea level
Acclimatizing
If the unit has been stored or transported at temperatures below +10 °C / +50 °F or above +40 °C / +104 °F, the unit will
require approx. 3
hours to acclimatize at room temperature.
Standards
Classification according to EC Directive 93/42/EEC II b
Classification according to Regulation (EU) 2017/745 II b
Type as per EN 60 601-1 Defibrillation-proof type CF applied part