Essilor E-Tess Tracer
6. ERROR CODES
The actions required if an error occurs are indicated below.
If one of these codes appears repeatedly or remains after the actions proposed below, record this code
and contact your technician.
Opt1 Contact your technician with the machine data.
Opt1000
An error occurred during startup.
Restart the device.
Opt4000
Opt4001
Opt4002
Opt4003
Opt4004
Opt4005
Opt4006
Opt4007
An error occurred during operation.
Restart the machine.
If the error occurs several times, contact your technician.
Opt6000
Opt6001
This function is not implemented.
Contact your technician or perform another action.
Opt6002
Insert a pattern or a frame in the tracing table.
Restart your cycle.
Opt6003
Opt6004
An error occurred during tracing.
Repeat the operation.
If the error occurs several times, contact your technician.
Opt6005
There is an error on your tracing table.
Contact your technician.
Opt6006
Opt6007
Opt6008
The bar code read is not compatible with your tracer.
Check your bar code reader and your code.
Repeat the operation.
Opt6011 Insert new standards, see Chapter 6.2.
Opt6012 Contact your technician.
Opt6014 Calibrate your tracer, see Chapter 5.2.
Opt6017
The file format of your standards is incorrect,
Delete your standard files from your tracer, see Chapter 2.2.
Insert new standards, see Chapter 6.2.
Opt6019
An error occurred on OMA file.
Restart the device and repeat the operation.
Opt6021
Opt6022
Contact your technician.
Opt6023
An error occurred during the software update.
Restart the device, the instrument will automatically start up on the old version.
Opt6024
Opt6025
The external USB stick is missing.
Insert the USB stick and continue.
Opt6026
There is an error on the update file.
You can use your tracer with your old version.
Contact your technician.
Opt6027 The pattern standard is incorrectly positioned, see Chapter 4.3., special cycles.
Opt6030 Your edger cannot carry out this type of machining (small B-dimension).
Opt6031 Your edger cannot carry out this type of machining (high curve).
Opt6032
The frame standard is incorrect or badly positioned.
Refer to Chapter 5.1. and repeat the operation.
Opt6033
There is an object in the tracing table.
Check your tracing table.
Restart the device and repeat the operation.
Opt6035 Insert the pattern standard supplied with your tracer, see Chapter 5.
Opt6036 Insert the frame standard supplied with your tracer, see Chapter 5.