Number of Foam Rinse Cycles: Enter the number of additional sparge tube
rinse cycles when a foaming sample is detected.
NOTE: Operational with EST Encon Evolution concentrator only.
Concentrator #1 Cycle Time (min): time of the total P&T cycle for
Concentrator #1. (Time from Purge Ready to Purge Ready.)
Concentrator #2 Cycle Time (min): time of the total P&T cycle time for
Concentrator #2. (Time from Purge Ready to Purge Ready.)
NOTE: When configured in dual concentrator mode double the total GC run
time to GC ready time and enter this value for both Concentrator #1 and
Concentrator #2.
Water Heater Temperature (°C): Enter the rinse water temperature.
Recommended temperature 85
o
C.
Syringe General Setup (Figure 3v):
Sample Type: Select from Blank, Soil or Water by <clicking> through the field.
Sample Fill Mode: Syringe
Sample Volume (ml): Enter the volume required, 5ml to 25ml in 1ml increments
Sample Prime Time (sec): Enter the time to prime the sample to the syringe.
Syringe Rinse Volume (ml): Enter yes or no to rinse the syringe and the volume
of the rinse needed. (5ml to 25ml in 1ml increments)
Number of Syringe Rinse Cycles: Enter the number of times the syringe should
be rinsed.
NOTE: See Loop General Setup for remainder of parameter explanations.
3.10.2 Internal Standards: <click> on the Internal Standards tab and the screen will
appear as shown in Figure 3w.