Figure 4ss: Gripper Axis Pop-up
28. <click> OK and the Gripper axis can be calibrated. See Figure 4qq.
29. Adjust the vertical height of the gripper so that there is at least a 1 to 2mm space
between the top of the vial and the top of the gripper. Refer to step 10.
30. When this is adjusted <click> Update Calibration. See Figure 4tt for image of
gripper axis.
Figure 4tt: Image of Space Between the Gripper and the Vial
31. Remove the calibration block from the wash station. Note: This is a user prompt in
the software, see Figure 4uu.
Figure 4uu: Prompt to remove the calibration block.
32. The next calibration step involves ensuring the vial has proper clearance in order for
the vial to be deposited into the soil cup. See Figure 4vv