EasyManua.ls Logo

EST Centurion WS - Page 87

Default Icon
205 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.6.4 Heater Calibration: <click> Heater Calibration and a popup will be displayed as
shown in Figure 4al. NOTE: This is Service ONLY Access. The PID settings should
NEVER be changed unless advised by EST Analytical Technical Support.
Figure 4al: Heater Calibration Menu
4.6.5 Digital Input / Output - <click> the Digital Input/Output selection and a screen will be
displayed as shown in Figure 4am. <click> on any of the input/output functions and the
selected function will be activated or de-activated. When activated, the function is
highlighted green. When de-activated, the function is highlighted blue.
Figure 4am: Digital Input/Output Menu
4.6.6 Internal Standard Calibration: <click> the Internal Standard selection and a screen
will be displayed as shown in Figure 4an.
1. Prepare a 40mL vial containing 1µL (or µl amount of your choice) of internal
standard for every 5mL of de-ionized water.

Related product manuals