FEASA LED ANALYSER
FUNCTIONAL VERSION 
Capture Mode
CAPTUREMULTI  - Multiple Capture Command
Transmit Receive
setrange@@#
putrange@@#
getrange@@
capturemulti
OK
OK
1 to 5
OK
Where:
@@            represents the fiber number in the range 01 – 20.
#               represents the capture range 1 - 5
Description
The  capturemulti  or  cm  command   will capture   the  LED  data  for   the  specified   fiber  at  the
specified range. 
The setrange@@# or putrange@@# commands will set the required range for each declared
individual fiber. The getrange@@ will return the range the user has enables for this fiber.
Example:
The DUT has 5 leds with 5 very different intensity values. You can now specify fiber 1 to range
Ultra (C5), Fiber 2 to range Super (C4), Fiber 3 to range High (C3), Fiber 4 to range Medium
(C2)  and  Fiber  5  to  range  Low  (C1).  Do  one  capturemulti  or  CM  command  to  get  your 5
readings.
Putrange015 (Fiber 1, Range Ultra / 5)
putrange024 (Fiber 2, Range Super / 4)
putrange033 (Fiber 3, Range High / 3)
putrange042 (Fiber 4, Range Medium / 2)
putrange051 (Fiber 5, Range Low / 1)
capturemulti
Back to Index
23