FEASA LED ANALYSER
FUNCTIONAL VERSION 
User Calibration Mode
getXOFFSET## - Return the x Chromaticity offset
Transmit Receive
getxoffset## ±0.xxx 
Where:
##          represents the Fiber Number and is a number in the range 01 – 20.
±0.xxx   represents the x Chromaticity offset
Description
This command is used to return the x Chromaticity offset for the LED under the Fiber number
##.   The value of this offset must be set by the setxoffset## command.    The default value
is 0.000.
Example:
The   PC   transmits  getxoffset01  to   the   LED   Analyser   to   instruct   it   to   send   the   stored   x
Chromaticity offset for the LED positioned under Fiber No 1.   The LED Analyser will return a
string  ±0.xxx to the PC.
getxoffset01
+0.155
Back to Index
62