FEASA LED ANALYSER
FUNCTIONAL VERSION 
User Calibration Mode
getWAVELENGTHOFFSET@@ - Get the Dominant Wavelength Offset
Transmit Receive
getwavelengthoffset@@ +/-XX
Where:
@@         represents the Fiber Number and is a number in the range 1 – 20.
+/- xx    represents the wavelength offset set of the LED in nanometers.
Description
This   command  is  used   to   get  the   value   of   the   Dominant   Wavelength   Offset   set  for   Fiber
number @@. 
The range of value that can be programmed is +/-99, however setting a value of greater than
+/-10 would probably indicate an error in your measurement setup   
Example:
The  PC   transmits  getwavelengthoffset01  to   the LED   Analyser  to  instruct   it  to  send  the
programmed dominant wavelength offset for the LED positioned under Fiber No 1.   The LED
Analyser will return a string  +/- xx to the PC.
getwavelengthoffset01
+07
Back to Index
65