User Calibration Mode
getCALIBRATIONset -  Select the Calibration Set to use
Transmit Receive
getcalibrationset ##
Where:
##             represents the set value (## can be 00 to 50).
Description
This   command   will   retrieve   which   calibration   set  you are accessing.    The   Led   analyser   is
capable of storing fifty one (51) separate user Calibration sets on the board.
The command  SETCALIBRATION##  stores the Calibration sets both on RAM and on EEPROM
(stays after power cycle but limits the number of set command operations to ~ 100K). 
The command  PUTCALIBRATION##  only  stores  them on  RAM and  the analyser  will default
back to what values were stored in the EEPROM after a power cycle.
The default calibration set is 00
Example:
If you want to know what set is active
getcalibrationset
03
Back to Index