Automatically accumulation is activated when Auto Accumulation is set to On in Auto 1, Auto 2, Auto 3 or Manual
mode is selected in F16 and/or F17.
Under the automatic accumulation mode, corresponding results will be accumulated automatically.
6.7.2 Manual Accumulation
Manual Accumulation is activated when scanner, PC or CMD mode is selected for both F16 and/or F17.
Under the Manual Accumulation mode, press [Print/M+] to accumulate the current value to memory.
6.7.3 When data is accumulated to memory
1. When a result is accumulated to memory, this instrument displays “n____x”. M+ Indicator appears to
indicate that memory contains stored data. “x” means the total number of transactions accumulated to
memory.
2. This instrument returns to normal display status after 2 seconds.
6.7.4 Memory recall and clearance
Accumulation data will be stored in memory and will not be erased by normal power off (by pressing the [ON/Off]
key) process. But changing weight unit or to another working mode will automatically erase accumulation data
stored.
Follow below procedures to recall and clear accumulation data.
1. Press [MR] to recall total accumulated weight.
2. Instrument flashes between “A____Y” (Y means the number of transactions accumulated) and total
accumulated result.
3. At this point: -
• Press [Zero] to quit, or
• Press [CE/x10] to clear memory. After [CE/x10] is pressed, instrument display Clear and M+
Indicator disappears to indicate all no data is stored in memory.
4. Press [MR] to recall total accumulated weight.
5. Instrument flashes between “A____Y” (Y means the number of transactions accumulated) and total
accumulated result.
6. At this point: -
• Press [Zero] to quit, or
• Press [CE/x10] to clear memory. After [CE/x10] is pressed, instrument display Clear and M+
Indicator disappears to indicate all no data is stored in memory.
6.8 Extended Display Mode
When F68 is set to OIML or NTEP, by pressing [CE/x10] the weighing resolution will be temporary (for 5 seconds)
changed to 10 times finer. Display keeps flashing when instrument is displaying the extended result.