Offset out of range / unstable during power on (5 minutes for OIML and NTEP mode)
Logic error. HI limit set is lower than LO limit (and HI is not = 0)
Logic error. LO limit is higher than HI limit (and HI is not = 0)
(F25 Mode 3) Container weight is higher than tArE H
(F25 Mode 3) Container weight is lower than tArE L
(F25 Mode 3 and Mode 4) Actual weight is higher than SP3
Exceed maximum power on (5 minutes for OIML and NTEP mode)
Capacity or division setting error (Division set is higher than 10000d)
Manual Zero and Tare stability error
Capacity setting error, Capacity 1 > Capacity 2
Division setting error, e1 > e2
Not able to obtain stable status for longer than 10 sec
Theoretical calculated value per e of Cal 2 varies more than 1% as of Cal 1.
Properly a load cell problem.
Mass value of Cal 2 is less than 150% of Cal 1.
Maximum accumulation limit is exceeded.
Standard deviation data exceed memory size (300 transactions).
Overload (Gross weight is more than Max plus 9d)
Major system error detected. Power off instrument and remove power adaptor
immediately. Then check load cell connection and system power status.
Negative Weight values exceeds display range
Important parameters have been changed. Power off and then power on instrument
again to reboot.
Negative Tare value exceeds display range
19. Daily Care & Maintenance
a. Clean the instrument with a soft, damp cloth. If necessary, use a mild detergent in water.
b. Do not use any harsh, abrasive material, acetone, volatile solvent, thinner or alcohol for cleaning.
c. Verify the accuracy of this instrument periodically. Re-calibrate if necessary. In some countries, calibration
requires authorized/qualified agent. Contact your dealer for more information.
d. Store this instrument in a dry and clean place,
e. Recharge battery before and every 4 months during long time storage.