50
Appendix B2: - Custom PC Output Protocol
Under this mode: -
• 23 different transaction data,
• 7 control codes, and
• 2 data separation types.
are available from instrument. Setup procedures: -
1. Go to F16 or F17 depends on output comport number.
2. Select PC, then set baud rate, parity and data length.
3. Select Custom in Protocol page. Then press [Print/M+].
4. Instrument displays Separa then followed by data separator selection page.
Notes: -
• Data separator is a symbol inserted between transaction data and is usually used by computer program to
separate various data.
• No data separator is inserted between in front or after control commands.
5. Select preferred data separator CoMMA (comma) or SemiCo (semi colon) then press [Print/M+].
6. Instruments displays Item number, then followed by the content page.
Note: - Item number means output sequence, e.g. Item 1 = the first content to output, item 3 = the third content to output.
7. Press [Func] and [Unit] until the preferred content appears. then press [Print/M+]. Refer to below Custom
PC Output Content Table for details.
8. Repeat step 6 to include other transaction data or control.
9. To complete and save a Custom output, select End then press [Print/M+].
10. Then select output time interval (refer to 5.5.1 for details).
11. Press [Print/M+] to save.