Appendix B2: - Custom PC Output Protocol
Under this mode: -
23 different transaction data,
7 control codes, and
2 data separation types.
are available from instrument. Setup procedures: -
1. Go to F16 or F17 depends on output comport number.
2. Select PC, then set baud rate, parity and data length.
3. Select Custom in Protocol page. Then press [Print/M+].
4. Instrument displays Separa then followed by data separator selection page.
Notes: -
Data separator is a symbol inserted between transaction data and is usually used by computer program to
separate various data.
No data separator is inserted between in front or after control commands.
5. Select preferred data separator CoMMA (comma) or SemiCo (semi colon) then press [Print/M+].
6. Instruments displays Item number, then followed by the content page.
Note: - Item number means output sequence, e.g. Item 1 = the first content to output, item 3 = the third content to output.
7. Press [Func] and [Unit] until the preferred content appears. then press [Print/M+]. Refer to below Custom
PC Output Content Table for details.
8. Repeat step 6 to include other transaction data or control.
9. To complete and save a Custom output, select End then press [Print/M+].
10. Then select output time interval (refer to 5.5.1 for details).
11. Press [Print/M+] to save.