5.4 Internal Function Table
Refer to the below tables for internal function number, parameter and setting notes.
Parameters / Note
Default = **
Internal Analogue to Digital (ad) Value
Press [Print/M+] to set offset value to zero when unloaded. Then add load on the platform to observe the span value of load applied.
When ADC is more than 1 million, LO Light Bar will appear. Actual ADC is = 1 million plus the ADC value being displayed.
When ADC is more than 2 million, OK Light Bar will appear. Actual ADC is = 2 million plus the ADC value being displayed.
When ADC is more than 3 million, HI Light Bar will appear. Actual ADC is = 3 million plus the ADC value being displayed.
Press [Zero] to quit to F1.
All Segment & Battery Voltage Check
All display segments and capacity bars will be lit on. Check any segments or LED of bars are missing.
To check battery voltage, press [Check].
Capacity, Division & Default Weight Unit
Display basic metrology characteristics (capacity, division and weight unit). Value displayed = Max + 1e
Press [Print/M+] to check current date value. To change date value, press [Print/M+] then enter date value and then press [Print/M+] to confirm.
To change time, press [Print/M+], then enter a new value and press [Print/M+].
System Initialization
(Set F7~F32) to Default)
If YES is selected, press [1] when “SURE?” is displayed. Indicator shows Done when initialization is completed.
Auto Power Off Time (Minute)
Auto power off function will be disabled when an energized power adaptor is plugged in.
1 = dimmest; 5 = brightest
Press [Func] or [Unit] to change setting and then press [Print/M+].
Weight Unit Enable / Disable
Select: -
1 (strongest filter) for bad working environment where vibration, wind flow… etc. affect stable reading,
5 for normal environment,
9 (least filter) for very good working environment where wind and vibration have no effect to stable reading.
Press [Zero] to quit, or press [Set] to select conversion speed.
4 parameters are available. Press [Func] or [Unit] to select: -