EasyManuals Logo

Fisher FIELDVUE DVC2000 User Manual

Fisher FIELDVUE DVC2000
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
Instruction Manual
D103176X012
Detailed Setup and Calibration
June 2017
24
Self Test Failures for Instrument Shutdown
Field Communicator Setup & Diag > Detailed Setup > Self Test Shutdown (1-2-8)
Upon shutdown, the instrument attempts to drive its output pressure to the zero current condition and no longer
executes its control function. In addition, the appropriate failure statuses are set. Once the problem that caused the
shutdown has been fixed, the instrument can be restarted by cycling the power or selecting Restart from the Mode
menu of the Field Communicator. Also see the Viewing Instrument Status section on page 35 for further details about
failures.
Follow the prompts on the Field Communicator display to determine the self test shutdown criteria from the
following:
DDone—Select this if you are done modifying the self test shutdown criteria.
DFlash ROM Fail—When enabled, the instrument shuts down whenever there is a failure associated with flash ROM
(read only memory).
DTemp Comp Fail—When enabled, the instrument shuts down whenever this is a failure associated with Temperature
Compensation.
DRef Voltage Fail—When enabled, the instrument shuts down whenever there is a failure associated with the internal
voltage reference.
DDrive Current Fail—When enabled, the instrument shuts down whenever the drive current does not read as expected.
DNVM Fail—When enabled, the instrument shuts down whenever there is a failure associated with NVM (non-volatile
memory).
DTemp Sensor Fail—When enabled, the instrument shuts down whenever there is a failure associated with the internal
temperature sensor.
DPress Sensor Fail—When enabled, the instrument shuts down whenever there is a failure associated with the pressure
sensor.
DTravel Sensor Fail—When enabled, the instrument shuts down whenever there is a failure associated with the travel
sensor.

Table of Contents

Other manuals for Fisher FIELDVUE DVC2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fisher FIELDVUE DVC2000 and is the answer not in the manual?

Fisher FIELDVUE DVC2000 Specifications

General IconGeneral
ManufacturerFisher
CategoryController
Device TypeDigital Valve Controller
Communication ProtocolHART
Input Signal4-20 mA
Enclosure RatingNEMA 4X, IP66
Operating Temperature-40 to 85°C (-40 to 185°F)
MaterialAluminum
Power SupplyLoop Powered
DiagnosticsAdvanced diagnostics
Output Signal4-20 mA
CertificationsATEX, IECEx, CSA, FM
MountingIntegral mounting to actuator

Related product manuals