Chapter 4: Maintenance
Periodic Maintenance
48 Hyperion Imaging System: User Guide
2 Scrub the parts with the recommended glassware brush (PN 20108, Restek).
3 Rinse thoroughly with DIW.
4 Spray isopropanol on the torch and air-dry.
5 Alternatively, completely dry the glassware with a blow dryer or heat gun.
NOTE Ensure that the parts are completely dry before reinstalling.
69BCoupling Tube
IMPORTANT We do not recommend cleaning of the coupling tube. Inspect the tube
periodically to ensure that there are no breaks or cracks. It is recommended to apply
compressed air on the connectors of the coupling tube to remove the buildup of dust
periodically. Replace the Coupling Tube Kit (PN 108480) if necessary.
Periodic Maintenance
70BVacuum Interface Cones
The vacuum interface cones are inspected and cleaned if necessary during the preventive
maintenance of the instrument. If running solution mode, clean the skimmer-reducer and
sampler cone weekly. See the Helios User Guide (PN 400250) for the procedure to remove,
cleaning, and reinstall the Vacuum Interface Cones.
71BReplace the Load Coil
If the load coil (PN 105398) is damaged or misshapen it must be replaced. Inspect the load
coil regularly to ensure that it is in good condition.