1652C/1653B/1654B
Users Manual
12
Table 6. Display Features (cont.)
No. Annunciator Meaning
=
Indicates the preset fault voltage limit.
The default setting is 50 V. Some
locations require the fault voltage be set
to 25 V, as specified by local electrical
codes.
Press when you turn on the tester to
toggle the fault voltage between 25 V
and 50 V. The value you set will appear
on the display and will be saved when
you turn the tester off.
Primary display and measurement units.
Memory locations. See page 37 for
detailed information on using memory
locations.
Low battery icon. See “Testing and
Replacing the Batteries” on page 41 for
additional information on batteries and
power management.
Appears when you press the Recall
button and you are looking at stored
data.
Appears when you press the Memory
button.
Appears when you press the Test
button. Disappears when the test is
completed.