Parameter Configuration
Logging Menu
If you call the logging menu, then you can select between two further
submenus: Menu for adjusting the averaging time and for adjusting the event
threshold in recording adjustments.
In the menu Averaging you select the time over which the data are to be
averaged. You can also select these values from the predefined values for
averaging time. As you change the averaging time, you will see on the display
the resulting logging time available for each averaging interval.
edx009.bmp
With Custom Setting you can choose any averaging time value. Depending on
the selected averaging time, maximum available recording time is indicated in
the display at the same time. Up to 4,320 averaging intervals can be recorded
with the record function.
By selecting the menu Event Thresholds you can select the threshold voltage at
which the recording is to be started (see also ‘Harmonics’).