1735
Users Manual
28
Logging Function
With Record/Measure you can start the logging function or you can get back
into the measuring mode. Before the start, the maximum logging time is
indicated and you can change this value with Esc followed by entering using
the Cursor key.
Changes in the averaging time, cause corresponding changes in the logging
time of the measurement (double averaging time = double logging time).
When the logger graphics reach the screen margin during recording, a picture
of this screen is saved.
The display is then deleted and the recording is continued. Up to 6 auto screens
are saved in the course of a recording. The saved screenshots can be retrieved
via the View Auto Screenshots menu.
Note
Do not forget to operate the Logger with the AC adapter during
logging in order to prevent shutdown caused by low battery.
edx025.bmp
21 Select between the individual phases
Select between the two representation modes:
- V and I (see fig.)
4 5
- V and F
- V and In