EasyManua.ls Logo

Fluke 732B - Replacing the Internal Battery

Fluke 732B
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
Powering the Standard
3
3-13
2. Verify that the BAT switches on both the 732B and 732B-7001 are set to l.
3. Charge the 732B and 732B-7001 for at least 36 hours.
4. Switch both BAT switches to
.
5. Fabricate and connect a cable with a 9-pin connector on each end so that the battery
output from the 732B-7001 is connected to the external dc input of the 732B. Refer
to Figures 3-7 and 3-8 for the connector pinouts.
6. Set the 732B-7001 BAT switch to l to enable its output.
7. Set the 732B BAT switch to l to connect the batteries in parallel.
8. After the ac line power to both the 732V and the 732B-7001 is turned of, the 732B
can be powered by battery for at least 130 hours.
Replacing the Internal Battery 3-13.
If the battery fails to switch from constant-current charging mode to float-charging
mode, either the charger is defective or the battery needs replacement. For preventive
maintenance, it is recommended that you have the battery replaced every 18-24 months
to ensure 72-hour battery backup time. Refer to Section 5 for how to obtain the correct
battery type and for the procedure to replace the battery.
1
2
345
9
8
7
6
DESCRIPTION
EXT BAT COMMON
EXT COMMON FOR IN CAL
EXT +5 VDC FOR IN CAL
OVEN TEMPERATURE THERMISTOR A
EXT BAT +
EXT BAT COMMON
EXT IN CAL SIGNAL (ASSERTED HIGH)
OVEN TEMPERATURE THERMISTOR B
EXT BAT +
PIN
NUMBER
1
2
3
4
5
6
7
8
9
k11f.eps
Figure 3-7. 732B MONITOR/EXT BAT IN Connector Pinout
www.valuetronics.com

Table of Contents

Related product manuals