EasyManua.ls Logo

Fluke 741B - Specifications

Fluke 741B
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
741B/743B
Users Manual
102
80T-150U Temperature Probe
80PK series thermocouples
80i-410 Clamp-on DC/AC Current Probe
80i-1010 Clamp-on DC/AC Current Probe
80i-500s Clamp-on AC Current Probe (requires
the Y9108 adapter)
80i-1000s Clamp-on AC Current Probe (requires
the Y9108 adapter)
80i-kW Current and Power Probe
Specifications
All specifications apply from +18
°
C to +28
°
C unless
stated otherwise.
All specifications assume a 5 minute warmup period.
Measurement specifications are valid only when
Damping is turned on. When damping is turned off,
or when the
g annunciator is displayed, floor
specifications are multiplied by 3. Floor specifications
are the second part of the specifications, usually
expressed as "% of full scale." The measure
pressure, temperature, and frequency functions are
specified only with damping on.
The standard specification intervals for the 741B and
743B are 1 and 2 years. Typical 90-day source and
measurement accuracy can be estimated by dividing
the 1 year "% of Reading" or "% of Output"
specifications by 2. Floor specifications, expressed
as "% of f.s.", remain constant.
To achieve the best noise rejection, use battery
power and tie all three common jacks together.

Table of Contents

Other manuals for Fluke 741B

Related product manuals