EasyManua.ls Logo

Fluke 741B - Test Comments; Calibrating a Delta-Pressure Flow Instrument

Fluke 741B
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Documenting Process Calibrator
Calibrating a Process Instrument
73
Test Comments
Model 743B runs tasks (custom procedures) that are
developed using a host computer and compatible
application software. A task may display a list of
proposed comments during execution. When the
comment list is displayed, select a comment to be
saved with the test results by pressing the u and d
keys followed by e.
Calibrating a Delta-Pressure Flow
Instrument
The procedure to calibrate a
instrument is the
same as for other instruments, as just described,
with the following differences:
Source square-root is automatically enabled
after the
As Found calibration template is
complete.
Measure/Source displays are in engineering
units.
The measurement percentage is automatically
corrected for the transmitter’s square-root
response, and is used to compute instrument
errors.
You select the
instrument procedure in a menu
after you press the
As Found softkey.

Table of Contents

Other manuals for Fluke 741B

Related product manuals