Documenting Process Calibrator
Using Measure Mode
41
Measurements in Percent of Scale
This feature lets you scale the measurements in
accordance with a particular process instrument’s
response. Percent of scale works for linear-output
transmitters or square-law transmitters such as
differential pressure transmitters that report flow rate.
Linear-Output Transmitters
1.
If necessary, press M for MEASURE mode.
2.
Select a measurement function (m, v, q,
t, or p) as previously described.
3.
Press the
Scale softkey.
4.
Use the numeric keypad to enter the 0% of scale
value (
0% Value).
5.
Press e.
6.
Use the numeric keypad to enter the 100% of
scale value (
100% Value).
7.
Press e.
8.
Leave
Mode set to %.
9.
Press the
Done softkey.
Percent of scale remains in effect until you press the
Scale softkey again or change to another
measurement function.
Square-Law Process Variables
When you select √ within scaling, the calibrator takes
the square root of its input and displays the
measurement in percent. For example, when
connected to the output of a delta-pressure
transmitter, the calibrator reading is proportional to
flow rate.
1.
If necessary, press M for MEASURE mode.
2.
Select a measurement function (m, v, q,
t, or p) as previously described.
3.
Press the
Scale softkey.
4.
Use the numeric keypad to enter the 0% of scale
value (
0% Value).