EasyManua.ls Logo

Fluke 741B - DC Current Measurement; Resistance Measurement

Fluke 741B
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Documenting Process Calibrator
Specifications
105
DC Current Measurement
Range Resolution
Models 741B and 743B
% of Reading +% of Full Scale
1 Year 2 Year
30 mA 1 µA 0.01% + 0.015% 0.02% + 0.015%
110 mA 10 µA 0.01% + 0.015% 0.02% + 0.015%
Temperature Coefficient: (0.001% of rdg. + 0.002% f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.01% f.s./(Common Mode Volt)
Maximum Input Voltage: 30V dc
Resistance Measurement
Range Resolution
Models 741B and 743B
% of Reading + ohms
1 Year 2 Year
11 0.001 0.05% + 0.05 0.075% + 0.05
110 0.01 0.05% + 0.05 0.075% + 0.05
1.1 k 0.1 0.05% + 0.5 0.075% + 0.5
11 k 1 0.1% + 10 0.1% + 10
Temperature Coefficient: (0.01% f.s. + 2 mΩ) /°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.005% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc

Table of Contents

Other manuals for Fluke 741B

Related product manuals