Documenting Process Calibrator
Specifications
117
Top and Bottom Limits of Ranges with Auto Range On
Range, dc V Measure Top of Range Bottom of Range
110 mV ±110.000 mV 0.000 mV
1.1V ±1.10000V ±0.10000V
11V ±11.0000V ±1.0000V
110V ±110.000V ±10.000V
300V ±300.00V ±100.00V
Range, dc V Source
110 mV +110.000 mV -10.000 mV
1.1V +1.10000V +0.10000V
11V +11.000V +1.1000V
Range, ohms Measure and Source
11Ω 11.000Ω 0.000Ω
110Ω 110.00Ω 10.00Ω
1.1 kΩ 1100.0Ω 100.0Ω
11 kΩ 11.000 kΩ 1.000 kΩ