EasyManua.ls Logo

Fluke 741B - Operating Features

Fluke 741B
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
741B/743B
Users Manual
12
Operating Features
Input and Output Jacks
Figure 5 shows the calibrator input and output jacks.
Table 2 explains their use.
Table 2. Input/Output Jacks and Connectors
No. Name Description
1 Battery Eliminator jack Jack for the Model BE9005 Battery Eliminator. Use the battery eliminator for bench-top
applications where ac line power is available. This input does not charge the battery.
2 wSERIAL PORT
(743B Only)
Connects the calibrator to an RS-232 serial port on a personal computer.
3 Pressure module
connector
Connects the calibrator to a pressure module.
4 TC input/output Jack for measuring or simulating thermocouples. This jack accepts a miniature
polarized thermocouple plug with flat, in-line blades spaced 7.9 mm (0.312 in) center to
center.
5, 6 w MEAS V jacks Input jacks for measuring voltage, frequency, or three- or four-wire RTDs (Resistance
Temperature Detectors).
7, 8 w SOURCE mA,
MEAS mA RTD
jacks
Jacks for sourcing or measuring current, measuring resistance and RTDs, and
supplying loop power.
9,10 w SOURCE V
RTD jacks
Output jacks for sourcing voltage, resistance, frequency, and for simulating RTDs.

Table of Contents

Other manuals for Fluke 741B

Related product manuals