EasyManua.ls Logo

Fluke 741B - Resistance Sourcing

Fluke 741B
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Documenting Process Calibrator
Specifications
109
Resistance Sourcing
Range Resolution
Models 741B and 743B
% of Output + ohms
1 Year 2 Year
11.000 1 m 0.01% + 0.02 0.02% + 0.02
110.00 10 m 0.01% + 0.04 0.02% + 0.04
1.1000 k 100 m 0.02% + 0. 5 0.03% + 0.5
11.000 k 1 0.03% + 5 0.04% + 5
Temperature Coefficient: (0.01% of f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C
Maximum and Minimum Current through Source Resistance:
11 Range: 3 mA dc max, 0.1 mA dc min
110 Range: 3 mA dc max, 0.1 mA dc min
1.1 k Range: 3 mA dc max, 0.01 mA dc min
11 k Range: 1 mA dc max, 0.01 mA dc min
Common Mode Error: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc