741B/743B
Users Manual
106
Continuity Testing
Tone Resistance
Continuous tone
May or may not get tone
No tone
<25
25 to 400
>400
Frequency Measurement
Ranges
Accuracy
1 Year 2 Year
1.00 Hz to 109.99 Hz 0.05 Hz 0.05 Hz
110.0 Hz to 1099.9 Hz 0.5 Hz 0.5 Hz
1.100 kHz to 10.999 kHz 0.005 kHz 0.005 kHz
11.00 kHz to 50.00 kHz 0.05 kHz 0.05 kHz
Minimum Amplitude for Frequency Measurement (square wave):
<1 kHz: 300 mV p-p
1 kHz to 30 kHz: 1.4 V p-p
>30 kHz: 2.8V p-p
Maximum input:
<1 kHz: 300V rms
>1 kHz: 30V rms
Input Impedance: 5 M
For frequency measurement less than 109.99 Hz, specifications apply for signals with a slew rate greater than 5
volt/millisecond.