741B/743B
Users Manual
108
DC Current Output
Range/Mode Resolution
Models 741B and 743B
% of Output + % of Full Scale
1 Year 2 Year
22 mA/ Source mA 1 µA 0.01% + 0.015% 0.02% + 0.015%
22 mA/ Simulate Transmitter
(Current Sink)
1 µA 0.02% + 0.03% 0.02% + 0.03%
Maximum Burden Voltage: 24V
Temperature Coefficient: (0.003% of output + 0.003% of f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc
Specification applies for currents between 2 mA and 22 mA. For current below 2 mA, typical accuracy is 0.15% of full
scale.