Documenting Process Calibrator
Using Source Mode
57
Sourcing in Percent of Scale
This feature lets you scale the output in accordance
with the input requirements of a particular process
instrument’s response. Percent of scale works for
linear-responding transmitters, or square-root
responding transmitters.
Linear-Responding Transmitters
1.
If necessary, press M for SOURCE mode.
2.
Select a source function (m, v, q, t,
or p) as previously described and enter a
value.
3.
Press the
Scale softkey.
4.
Use the numeric keypad to enter the 0% of scale
value (
0% Value).
5.
Press e.
6.
Use the numeric keypad to enter the 100% of
scale value(
100% Value).
7.
Leave
Mode set to %.
8.
Press the
Done softkey.
Percent of scale remains in effect until you press the
Scale softkey again or you change the source
function.
Square-Law Process Variables
When you select within scaling, the calibrator
output value is the percent value entered, squared,
and converted to engineering units.
1.
If necessary, press M for SOURCE mode.
2.
Select a source function (m, v, q, t,
or p) as previously described.
3.
Press the
Scale softkey.
4.
Use the numeric keypad to enter the 0% of scale
value (
0% Value).
5.
Press e.
6.
Use the numeric keypad to enter the 100% of
scale value(
100% Value).
7.
Set
Mode to .
8.
Press e.