EasyManua.ls Logo

Fluke 741B - Simultaneous Measure;Source

Fluke 741B
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Documenting Process Calibrator
Simultaneous Measure/Source
63
Simultaneous Measure/Source
Use the MEASURE/SOURCE mode to calibrate or
emulate a process instrument. Press M so that a
split screen display appears as shown below.
gj42s.eps
Table 7 shows the functions you can use
simultaneously when Loop Power is disabled. Table
8 shows the functions you can use simultaneously
when Loop Power is enabled.
You can use the
Step or Auto Step features to adjust
the output in MEASURE/SOURCE mode, or you can
use the calibration routine provided when you press
the
As Found softkey.
The following two softkeys displayed in
MEASURE/SOURCE mode are for use in calibrating
a process instrument.
As Found, which lets you set up a calibration
routine to obtain and record as found data.
Auto Step, which lets you set up the calibrator for
auto-stepping, as previously described.
Instructions for calibrating a process instrument
follow.

Table of Contents

Other manuals for Fluke 741B

Related product manuals