741B/743B
Users Manual
110
Frequency Sourcing
Range Accuracy
1 and 2 Year
0.00 Hz to 10.99 Hz 0.01 Hz
11.00 Hz to 109.99 Hz 0.1 Hz
110.0 Hz to 1099.9 Hz 0.1 Hz
1.100 kHz to 21.999 kHz 0.002 kHz
22.000 kHz to 50.000 kHz 0.005 kHz
Waveform Choices: Zero-symmetric sine or positive square wave, 50% duty cycle.
Amplitude: 0.1 to 10V pk
Amplitude Accuracy:
0 Hz to 1099 Hz: 3% of output + 0.5% f.s
1.1 kHz to 10.9 kHz: 10% of output + 0.5% f.s
11 kHz to 50 kHz: 30% of output + 0.5% f.s
Maximum Input Voltage: 30V dc