EasyManua.ls Logo

Fluke 741B - Page 77

Fluke 741B
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Documenting Process Calibrator
Calibrating a Process Instrument
67
8. Press
M for MEASURE/SOURCE mode. The
display changes to:
gj42s.eps
9. Press the As Found softkey, followed by the
Instrument softkey. The display changes to:
gj44s.eps
10. Enter values for 0% and 100% of 4.0 mA and
20.0 mA, respectively. Set
Tolerance to 0.5% of
span. (Use other values if necessary for your
application.)

Table of Contents

Other manuals for Fluke 741B

Related product manuals