Maintenance
INTERNAL CLEANING
6
6-71
6-78. Cleaning After Soldering
CAUTION
T.M.C. Cleaner and similar products can can attack the nylon
latches and other plastic pieces.
f6-21.wmf
Figure 6-21. Guard Crossing Test Waveforms
If a PCA has been soldered, it should first be cleaned with SPRAYON T.M.C Cleaner
(rosin flux remover) or equivalent. The PCA should then be cleaned with water as
described above.