EasyManua.ls Logo

Fluke 8842A - An Overview of Remote Operation; IEEE-488 Address Selection

Fluke 8842A
269 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8842A
Instruction Manual
3-4
A
D
D
R
E
S
S
T
A
L
K
O
N
L
Y
A5 A4 A3 A2 A1 A
D
D
R
E
S
S
T
A
L
K
O
N
L
Y
A5 A4 A3 A2 A1 A
D
D
R
E
S
S
T
A
L
K
O
N
L
Y
A5 A4 A3 A2 A1
00 0 0 0000 11 001011 22 0 10110
01 0 0 0001 12 001100 23 0 10111
02 0 0 0010 13 001101 24 0 11000
03 0 0 0011 14 001110 25 0 11001
04 0 0 0100 15 001111 26 0 11010
05 0 0 0101 16 010000 27 0 11011
06 0 0 0110 17 010001 28 0 11100
07 0 0 0111 18 010010 29 0 11101
08 0 0 1000 19 010011 30 0 11110
09 0 0 1001 20 010100 31 Not allowed
10 0 0 1010 21 010101 1 XXXXXTALK
ONLY
X = setting does not matter
f3-01.wmf
Figure 3-1. IEEE-488 Address Selection
Whenever the 8842A is in the local state, the IEEE-488 address can be displayed on the
front panel by pressing the LOCAL button.
3-4. AN OVERVIEW OF REMOTE OPERATION
An overview of remote operation is presented in the block diagram in Figure 3-2. Each
block represents a register, buffer, etc., contained in the 8842A. The status registers in the
center column indicate the instrument’s status, including its function, range, reading rate,
etc. The input buffer receives data from the IEEE-488 bus. The output buffer receives
data from the blocks to its left, and sends data on to the IEEE-488 bus.