SAFETY INSTRUCTIONS
1.5 IMPAIRED SAFETY
Whenever
H Is
likely
that safety has been Impaired, the Instrument must be turned off and
disconnected
from all external voltage sources, and the batteries must be removed. The matter
should then be referred to qualified technicians. Safety Is likely to be impaired if, for example,
the
instrument fails to perform the intended measurements or shows visible damage.
1 .6 GENERAL SAFETY INFORMATION
WARNING: Removing the instrument covers or removing parts,
except those to which
access can be gained by hand, is likely to expose live parts and
accessible
terminals which can be dang^ous to life.
The instrument must be disconnected frorr> all voltage sources and batteries must
be
removed before
rt is opened.
Capacitors inside the instrument can hold their charge even if
the
instrument
has been separated
from all voltage sources and batteries are removed. Components which are important
for
the safety
of the Instrument may only be replaced by components obtained through your local FLUKE/PH I LI PS
organization. These components are Indcated by an asterisk
(*)
In the parts list sechon (chapters).