EasyManua.ls Logo

Fluke 93 - Only

Fluke 93
267 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-22
PERFORMANCE
VERIFICATION PROCEDURE
7. Cursor measursmsnU:
time accuracy
*•*
Models 95/97 only!
This test checks the
accuracy of the cursors while measuring time.
Test equipment:
Tektronix TG 501 Time Mark Generator
Test setup:
SeoD^Meter
Setting/procedure:
A Apply a 1 ms time marker signal to charnel A. Use a 500 termination.
B Switch on the Scope Meter and press
the SCOPE key to get into SCOPE mode. Now press the
AUTO SET key.
C Set
the
timebase to 1 ms/div.
D Press
the HOLD/RUN key to freeze the display.
E Press
the CURSOR DATA key to get Into the
cursor menu,
F Press the CURSOR softkey to turn on the cursor
lines.
G Position the cursor lines with the
<CURSOR -l-> and <CUfiSOR
2->
keys, so that they cover
a
distance of 6 time marker Intervals.
Position the markers exactly lothe top of the marker
pulses.
See figure 4.11.
Requirements:
The measured
time distance between the cursors Is displayed
at the right side next to the traces. This
value
must be 5. 99... 6.01 ms.
*5VAC PROBE
*10
B OfF PROBE*10
Figure
4. 1 1 Cursor lines on marker pulses

Table of Contents

Related product manuals